Calibration
Calibration
Thermo Scientific
42iQD Instruction Manual
4-11
For detailed information about the menu parameters and the icons used in
these procedures, see the “
” chapter.
1.
Determine the GPT flow conditions required to meet the dynamic
parameter specifications as indicated in “
Specifications for Gas Titrator
” earlier in this chapter.
2.
Adjust the GPT diluent air and O
3
generator air flows to obtain the
flows determined in “
Dynamic Parameter Specifications for Gas
” earlier in this chapter. The total GPT airflow must exceed the
total demand of the analyzer. The 42iQD requires approximately 700
cc/min of sample flow, and a total GPT airflow of at least 1.5 liters/min
is recommended.
a.
Allow the analyzer to sample zero air until the NO
x
response
stabilizes.
b.
After the response is stabilized, from the Home Screen, choose
Calibration>Calibrate Background>Calibrate NO
x
Background
.
The Target NO
x
Concentration button will read 0.00. The
Calculated NO
x
Background button will display the background
needed to make the current NO
x
concentration go to 0.00.
c.
Press
Calibrate
to set the NO
x
reading to zero and to save the new
background.
d.
Record the stable zero air responses as Z
NOX
.
3.
Adjust the NO flow from the standard NO cylinder to generate a NO
concentration of about 80% of the upper range limit (URL) of the NO
range. The exact NO concentration is calculated from:
F
+
F
+
F
NO
x
F
=
]
[NO
D
O
NO
STD
NO
OUT
Where:
[NO]
OUT
= Diluted NO concentration at the output manifold, ppm
NO
STD
= No feed concentration
F
NO
= No flow
F
O
= Ozone flow
F
D
= Dilution flow
Содержание 42iQD
Страница 1: ...42iQD Instruction Manual NOx Analyzer 119013 00 1Sep2021 ...
Страница 2: ......
Страница 32: ...Operation Instrument Display 3 8 42iQD Instruction Manual Thermo Scientific ...
Страница 182: ...Maintenance Pump Rebuilding 5 8 42iQD Instruction Manual Thermo Scientific Figure 5 3 Pump Rebuilding ...
Страница 184: ......
Страница 198: ...Servicing Fan Replacement 7 8 42iQD Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 212: ...Servicing Power Supply Replacement 7 22 42iQD Instruction Manual Thermo Scientific Figure 7 26 Replacing Power Supply ...
Страница 235: ...Servicing DMC Ozonator Thermo Scientific 42iQD Instruction Manual 7 45 Figure 7 49 Flow Switch Remove ...
Страница 238: ......
Страница 272: ......
Страница 284: ......
Страница 285: ......