![Thermo Scientific 42iQD Скачать руководство пользователя страница 204](http://html.mh-extra.com/html/thermo-scientific/42iqd/42iqd_instruction-manual_1105143204.webp)
Servicing
I/O Replacement
7-14
42iQD Instruction Manual
Thermo Scientific
Use the following procedure to replace the I/O boards.
1.
Turn instrument OFF, unplug power cord, and remove the cover
2.
Swing arm open.
Figure 7–15.
I/O Replacement, Arm
3.
Pull board upwards.
Figure 7–16.
I/O Replacement, Remove Board
I/O Replacement
Содержание 42iQD
Страница 1: ...42iQD Instruction Manual NOx Analyzer 119013 00 1Sep2021 ...
Страница 2: ......
Страница 32: ...Operation Instrument Display 3 8 42iQD Instruction Manual Thermo Scientific ...
Страница 182: ...Maintenance Pump Rebuilding 5 8 42iQD Instruction Manual Thermo Scientific Figure 5 3 Pump Rebuilding ...
Страница 184: ......
Страница 198: ...Servicing Fan Replacement 7 8 42iQD Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 212: ...Servicing Power Supply Replacement 7 22 42iQD Instruction Manual Thermo Scientific Figure 7 26 Replacing Power Supply ...
Страница 235: ...Servicing DMC Ozonator Thermo Scientific 42iQD Instruction Manual 7 45 Figure 7 49 Flow Switch Remove ...