Servicing
DMC Converter Replacement
Thermo Scientific
42iQD Instruction Manual
7-37
Use the following procedure to replace the converter assembly board.
Equipment Required:
Phillips drive, #2
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover.
2.
Remove DMC Converter as described in “
” in this chapter.
3.
Unplug electrical J6 and J7.
4.
Using a #2 Phillips drive, unfasten four 6-32 pan head screws.
Figure 7–42.
Converter Assembly Remove bd
5.
Pull converter assembly board outwards.
6.
Replace converter assembly board and assemble in reverse order.
Converter Assembly
Board Replacement