Operation
Settings
3-82
42iQD Instruction Manual
Thermo Scientific
The Pressure Calibration screen is used to calibrate the pressure sensor to
zero, span, or factory default values.
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Pressure Calibration
The Pressure Calibration screen contains the following information:
●
Across:
●
Description:
Lists items in table.
●
Reading:
Displays reading of each pressure sensor.
●
Calibration:
Starts calibration or resets default values.
●
Down:
●
Sensor 1
−
3 Reading:
Under the column labeled Reading, current
reading of each pressure sensor.
●
Atmospheric Sensor 1
−
3:
Under the column labeled reading, the user
enters the current atmospheric pressure in mmHg units. Under the
column labelled Calibration, the user presses the Start button to
calibrate the high point of the sensor.
●
Zero Sensor 1
−
3:
User should put the pressure sensor under a strong
vacuum. Under the column labeled reading, the user enters the
pressure in mmHg. Under the column labeled Calibration, the user
presses the Start button to calibrate the low point of the sensor.
●
Reset all values:
Resets values to default.
Pressure Calibration
Содержание 42iQD
Страница 1: ...42iQD Instruction Manual NOx Analyzer 119013 00 1Sep2021 ...
Страница 2: ......
Страница 32: ...Operation Instrument Display 3 8 42iQD Instruction Manual Thermo Scientific ...
Страница 182: ...Maintenance Pump Rebuilding 5 8 42iQD Instruction Manual Thermo Scientific Figure 5 3 Pump Rebuilding ...
Страница 184: ......
Страница 198: ...Servicing Fan Replacement 7 8 42iQD Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 212: ...Servicing Power Supply Replacement 7 22 42iQD Instruction Manual Thermo Scientific Figure 7 26 Replacing Power Supply ...
Страница 235: ...Servicing DMC Ozonator Thermo Scientific 42iQD Instruction Manual 7 45 Figure 7 49 Flow Switch Remove ...
Страница 238: ......
Страница 272: ......
Страница 284: ......
Страница 285: ......