Operation
Settings
3-78
42iQD Instruction Manual
Thermo Scientific
The Reaction Chamber Settings screen allows the user to calibrate the
Input board and PMT.
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Reaction Chamber Settings
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Reaction Chamber Settings>Continue to Input Board
Calibration Screen
The Input Board Calibration screen contains the following information:
●
Across:
●
Description:
Defines the actions the user can do.
●
Input Board Cal:
User can manually adjust the input board value,
start/stop automatic calibration, read voltage, frequency and zero
offset frequency values, and reset default values.
●
Units:
Displays units for each item.
Reaction Chamber
Settings
Содержание 42iQD
Страница 1: ...42iQD Instruction Manual NOx Analyzer 119013 00 1Sep2021 ...
Страница 2: ......
Страница 32: ...Operation Instrument Display 3 8 42iQD Instruction Manual Thermo Scientific ...
Страница 182: ...Maintenance Pump Rebuilding 5 8 42iQD Instruction Manual Thermo Scientific Figure 5 3 Pump Rebuilding ...
Страница 184: ......
Страница 198: ...Servicing Fan Replacement 7 8 42iQD Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 212: ...Servicing Power Supply Replacement 7 22 42iQD Instruction Manual Thermo Scientific Figure 7 26 Replacing Power Supply ...
Страница 235: ...Servicing DMC Ozonator Thermo Scientific 42iQD Instruction Manual 7 45 Figure 7 49 Flow Switch Remove ...
Страница 238: ......
Страница 272: ......
Страница 284: ......
Страница 285: ......