Troubleshooting
Troubleshooting Guide
Thermo Scientific
42iQD Instruction Manual
6-5
Problem
Possible Cause
Action
Alarm – 5V/24V Step Board Cable connection
Check the cable connections to that
Step POL board.
Alarm – Analog I/O
Defective PCB
Replace Analog board.
Alarm – Digital I/O
Defective PCB
Replace Digital board.
Alarm – NO Bkg Check
Offset
Incorrect high alarm
limit
Verify the high limit is correct via
Settings>Status and
Alarms>Concentrations screen.
Instrument
background
calibration failed
Recalibrate the instrument.
Alarm – NO Span Check
Offset
Incorrect high alarm
limit
Verify the high limit is correct via
Settings>Status and
Alarms>Concentrations screen.
Instrument span
calibration failed
Recalibrate the instrument.
Alarm – Ambient
Thermistor
Defective
Flow/Pressure board
Replace defective DMC board as
needed.
Alarm – Chamber Pressure
Defective
Flow/Pressure board
Replace defective DMC board as
needed.
Alarm – Bench
Temperature Short
Cable connection
Check the cable connection from the
heater to the DMC board.
Broken wire
Verify the wires are properly
connected to both sides of the heater.
Defective heater
Replace heater as needed.
Defective DMC board Replace defective DMC board as
needed.
Alarm – Bench
Temperature Open
Cable connection
Check the cable connection from the
heater to the DMC board.
Broken wire
Verify the wires are properly
connected to both sides of the heater.
Defective heater
Replace heater as needed.
Defective DMC board Replace defective DMC board as
needed.
Alarm – Cooler Current too
Low
Defective cooler
Replace cooler as needed.
Defective DMC board Replace DMC board as needed.
Alarm – Cooler Current too
High
Defective cooler
Replace cooler as needed.
Defective DMC board Replace DMC board as needed.
Alarm – Cooler Voltage too
Low
Defective cooler
Replace cooler as needed.
Содержание 42iQD
Страница 1: ...42iQD Instruction Manual NOx Analyzer 119013 00 1Sep2021 ...
Страница 2: ......
Страница 32: ...Operation Instrument Display 3 8 42iQD Instruction Manual Thermo Scientific ...
Страница 182: ...Maintenance Pump Rebuilding 5 8 42iQD Instruction Manual Thermo Scientific Figure 5 3 Pump Rebuilding ...
Страница 184: ......
Страница 198: ...Servicing Fan Replacement 7 8 42iQD Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 212: ...Servicing Power Supply Replacement 7 22 42iQD Instruction Manual Thermo Scientific Figure 7 26 Replacing Power Supply ...
Страница 235: ...Servicing DMC Ozonator Thermo Scientific 42iQD Instruction Manual 7 45 Figure 7 49 Flow Switch Remove ...
Страница 238: ......
Страница 272: ......
Страница 284: ......
Страница 285: ......