Maintenance
Pump Rebuilding
Thermo Scientific
42iQD Instruction Manual
5-5
Use the following procedure to rebuild the pump.
Equipment required:
Pump rebuild kit
Phillips screwdriver, #1
Marking pencil or marker
Allen wrench, 2.5 mm
Figure 5–2.
Pump
Disassembly:
Note
Parts removed must be replaced exactly as found. If repairing
multiple pumps, take care not to mix parts.
▲
1.
Turn instrument OFF, unplug the power cord, and remove the cover.
Make a sketch of the position of any tubes and fittings for ease of re-
assembly later.
2.
Mark the position of the pump headplate, intermediate plate and
compressor housing relative to each other by drawing a line on the
edges with a pencil or marker to insure proper re-assembly.
3.
One of the aluminum compressor housing covers must be removed to
gain access to the inside of the compressor housing. Remove the four
Pump Rebuilding
Содержание 42iQD
Страница 1: ...42iQD Instruction Manual NOx Analyzer 119013 00 1Sep2021 ...
Страница 2: ......
Страница 32: ...Operation Instrument Display 3 8 42iQD Instruction Manual Thermo Scientific ...
Страница 182: ...Maintenance Pump Rebuilding 5 8 42iQD Instruction Manual Thermo Scientific Figure 5 3 Pump Rebuilding ...
Страница 184: ......
Страница 198: ...Servicing Fan Replacement 7 8 42iQD Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 212: ...Servicing Power Supply Replacement 7 22 42iQD Instruction Manual Thermo Scientific Figure 7 26 Replacing Power Supply ...
Страница 235: ...Servicing DMC Ozonator Thermo Scientific 42iQD Instruction Manual 7 45 Figure 7 49 Flow Switch Remove ...
Страница 238: ......
Страница 272: ......
Страница 284: ......
Страница 285: ......