Servicing
DMC Converter Replacement
7-36
42iQD Instruction Manual
Thermo Scientific
Use the following procedure to replace the converter.
Equipment Required:
Phillips drive, #2
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover.
2.
Unplug DMC cables 2X.
3.
Disconnect all plumbing 8X.
Note
The amount of connections will vary depending on the different
options installed.
▲
4.
Using a #2 Phillips drive, unfasten captive hardware 4X.
Figure 7–41.
Converter DMC Assembly Remove
DMC Converter
Replacement