Operation
Settings
3-46
42iQD Instruction Manual
Thermo Scientific
The NO
2
Converter screen provides status and alarms related to the NO
2
converter. If an item being monitored goes outside the lower or higher
alarm limit, an alarm is activated.
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Settings>Health Check>Status and Alarms>NO
2
Converter
The NO
2
Converter screen contains the following information:
●
Across:
●
Converter:
This column lists items associated with the NO
2
converter.
●
Value:
Displays the current value for each item.
●
Low Alarm:
Displays low alarm status for each item.
●
High Alarm:
Displays high alarm status for each item.
●
Units:
Displays units for each item.
●
Down:
●
NO
2
Converter Temperature:
Displays the current NO
2
converter
temperature reading. User can adjust low and high alarm limits.
●
NO
2
Thermocouple Open:
Displays OK/Fail for NO
2
thermocouple
open.
●
NO
2
Thermocouple Short:
Displays OK/Fail for NO
2
thermocouple
short.
●
Board Communication:
Displays OK/Fail for board communication
status.
NO
2
Converter
Содержание 42iQD
Страница 1: ...42iQD Instruction Manual NOx Analyzer 119013 00 1Sep2021 ...
Страница 2: ......
Страница 32: ...Operation Instrument Display 3 8 42iQD Instruction Manual Thermo Scientific ...
Страница 182: ...Maintenance Pump Rebuilding 5 8 42iQD Instruction Manual Thermo Scientific Figure 5 3 Pump Rebuilding ...
Страница 184: ......
Страница 198: ...Servicing Fan Replacement 7 8 42iQD Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 212: ...Servicing Power Supply Replacement 7 22 42iQD Instruction Manual Thermo Scientific Figure 7 26 Replacing Power Supply ...
Страница 235: ...Servicing DMC Ozonator Thermo Scientific 42iQD Instruction Manual 7 45 Figure 7 49 Flow Switch Remove ...
Страница 238: ......
Страница 272: ......
Страница 284: ......
Страница 285: ......