![Thermo Scientific 42iQD Скачать руководство пользователя страница 84](http://html.mh-extra.com/html/thermo-scientific/42iqd/42iqd_instruction-manual_1105143084.webp)
Operation
Settings
3-60
42iQD Instruction Manual
Thermo Scientific
The Preventive Maintenance screen reminds the user when certain
instrument components need to be serviced/replaced. When the “Months
Left” has decreased to 2, the row is highlighted yellow. If the “Months
Left” is 1 or less, the row is highlighted red and the maintenance icon
(gears) will appear in the status bar located at the bottom of the screen.
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Settings>Health Check>Maintenance>Preventive
Maintenance
The Preventive Maintenance screen contains the following information:
●
Across:
●
Component:
Device to be routinely serviced or replaced.
●
Interval in Months:
Expected period of time before a component
needs to be checked and/or serviced.
●
Months Left:
Count down timer in months. Remaining time since
the beginning of the service interval. When the value is 1 or less,
the row will be highlighted and it is suggested that the component
should be checked and/or serviced.
●
Reset:
Once the component is serviced/replaced, the user presses the
Reset button and the “Months Left” value resets to the “Interval in
Months” value.
●
Alert:
Allows the user to opt out of receiving preventive
maintenance alerts. Displays Enabled/Disabled for each
component.
Preventive Maintenance
Содержание 42iQD
Страница 1: ...42iQD Instruction Manual NOx Analyzer 119013 00 1Sep2021 ...
Страница 2: ......
Страница 32: ...Operation Instrument Display 3 8 42iQD Instruction Manual Thermo Scientific ...
Страница 182: ...Maintenance Pump Rebuilding 5 8 42iQD Instruction Manual Thermo Scientific Figure 5 3 Pump Rebuilding ...
Страница 184: ......
Страница 198: ...Servicing Fan Replacement 7 8 42iQD Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 212: ...Servicing Power Supply Replacement 7 22 42iQD Instruction Manual Thermo Scientific Figure 7 26 Replacing Power Supply ...
Страница 235: ...Servicing DMC Ozonator Thermo Scientific 42iQD Instruction Manual 7 45 Figure 7 49 Flow Switch Remove ...
Страница 238: ......
Страница 272: ......
Страница 284: ......
Страница 285: ......