Operation
Settings
3-74
42iQD Instruction Manual
Thermo Scientific
●
High Range NO
x
Setting:
Sets the high range NO
x
concentration range
for the analog output when in dual or auto range mode.
●
Low Range NO
x
Setting:
Sets the low range NO
x
concentration range
for the analog output when in dual or auto range mode.
Settable standard ranges according to unit selection include:
ppb
50
−
20000 ppb
ppm
0.05
−
20 ppm
%
0.000005
−
0.002 %
µg/m
3
100
−
30000 µg/m
3
mg/m
3
0.1
−
30 mg/m
3
g/m
3
.0001
−
0.030 g/m
3
Settable extended ranges according to unit selection include:
ppb
200
−
100000 ppb
ppm
0.2
−
200 ppm
%
00002
−
0.01 %
µg/m
3
500
−
150000 µg/m
3
mg/m
3
0.5
−
150 mg/m
3
g/m
3
.0005
−
0.150 g/m
3
Содержание 42iQD
Страница 1: ...42iQD Instruction Manual NOx Analyzer 119013 00 1Sep2021 ...
Страница 2: ......
Страница 32: ...Operation Instrument Display 3 8 42iQD Instruction Manual Thermo Scientific ...
Страница 182: ...Maintenance Pump Rebuilding 5 8 42iQD Instruction Manual Thermo Scientific Figure 5 3 Pump Rebuilding ...
Страница 184: ......
Страница 198: ...Servicing Fan Replacement 7 8 42iQD Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 212: ...Servicing Power Supply Replacement 7 22 42iQD Instruction Manual Thermo Scientific Figure 7 26 Replacing Power Supply ...
Страница 235: ...Servicing DMC Ozonator Thermo Scientific 42iQD Instruction Manual 7 45 Figure 7 49 Flow Switch Remove ...
Страница 238: ......
Страница 272: ......
Страница 284: ......
Страница 285: ......