Servicing
DMC PMT Cooler and Reaction Chamber Replacement
7-26
42iQD Instruction Manual
Thermo Scientific
Use the following procedure to replace the PMT cooler and clean or
replace the reaction chamber assembly.
●
Remove the PMT Cooler
●
Replacing the PMT board
●
Removing and Cleaning the Reaction chamber
●
Replacing the PMT Cooler and Reaction chamber
●
PMT Replacement
Use the following procedure to remove the PMT cooler from the
instrument.
Equipment Required:
Phillips drive, #2
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover (
).
Make sure the PMT has cooled down before touching.
▲
2.
Unhook plumbing by pulling loop downwards and unhook tubing.
3.
Using a #2 Phillips drive, unfasten two 6-32 flat head screws.
4.
Pull shroud outwards.
DMC PMT Cooler
and Reaction
Chamber
Replacement
PMT Cooler Shroud
Removal
Icon Here
Содержание 42iQD
Страница 1: ...42iQD Instruction Manual NOx Analyzer 119013 00 1Sep2021 ...
Страница 2: ......
Страница 32: ...Operation Instrument Display 3 8 42iQD Instruction Manual Thermo Scientific ...
Страница 182: ...Maintenance Pump Rebuilding 5 8 42iQD Instruction Manual Thermo Scientific Figure 5 3 Pump Rebuilding ...
Страница 184: ......
Страница 198: ...Servicing Fan Replacement 7 8 42iQD Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 212: ...Servicing Power Supply Replacement 7 22 42iQD Instruction Manual Thermo Scientific Figure 7 26 Replacing Power Supply ...
Страница 235: ...Servicing DMC Ozonator Thermo Scientific 42iQD Instruction Manual 7 45 Figure 7 49 Flow Switch Remove ...
Страница 238: ......
Страница 272: ......
Страница 284: ......
Страница 285: ......