Symptom
Using excessive anaerobic gas
Cause/Solution
Leak in the vinyl front (sleeves, cuffs, or gloves). Repair leaks as needed.
Symptom
Excessive moisture in the chamber.
Cause/Solution
Wafers need to be baked more frequently or replaced. Bake for 2 hours
at 160°C. May need to replace desiccant wafer.
Symptom
Vacuum pump runs continuously, but does not register on
gauge.
Cause/Solution
Leak on outer door seal. Adjust or replace door seal.
Symptom
Vacuum pump runs for a short time, then shuts down (gas sys-
tem light out).
Cause/Solution
Inner door seal is leaking. Ensure inner door is pulled fully forward and
latched. If unsuccessful, door seal or latch bar may require replacement.
Symptom
Anaerobic gas is continuously injected into glove chamber.
Cause/Solution
Obstruction in the anaerobic gas solenoid, defective gas pressure switch,
or manual purge valve is on.
Symptom
After completion of a cycle, the sequence incomplete light stays
on.
Cause/Solution
The inner/outer door switch is defective or needs adjusted.