Pull a vacuum on the interchange chamber (about 5” vacuum). Remove
the Allen screws from the hinge blocks and put the thread lock on the
threads of the screws. Reinstall the screws in the hinge blocks. Turn the
screw in a clockwise direction until the screws just touch the hinge pins of
the door. Leave the screws at this position.
Lift the door latch and turn the latch clockwise to adjust the inner door of
the interchange chamber. The latch is adjusted correctly when the latch
pulls down firm. When the latch is pulled down, the door will not move
from front to back.
Note
If the latch is turned fully clockwise and the door still moves, the
latch bar may need to be replaced (see Figure 3).
s
Anaerobic System 5-3
Thermo Scientific
Section 5
Troubleshooting
Outer Door Hinge
Alignment
Inner Door Latch
Adjustment
Loosen this screw
to tighten latch
20068 - #8-32 x 5/16
SS FPH UC Screw
1025-08-4
Ext Door Shim
Figure 1
1025-06-2 Stub Lever Latch Assembly
5
Stub Lever Latch
Lever Latch
(Or) Washers
Outer Door
Latch Shim
Figure 2
Adding Shims or Washers