Caution
The inner and outer interchange doors should not be open at the
same time while the work chamber is operating under anaerobic
conditions.
s
1. With the inner door securely closed and clamped, open the outer door
and place the material to be transferred into the interchange.
2. Secure the outer door.
3. For an automatic transfer cycle, press the Cycle Start button.
4. When the cycle is complete, the Equalize and both door lights go out,
and the Cycle Complete light comes on. Then it is safe to open the
inner door and remove the material.
5. After removing the material, be sure to secure the inner door.
Caution
Always keep the inner door securely closed to prevent atmospheric
contamination due to accidental outer door openings. The outer door
should also be kept closed at all times.
s
To remove the material from the chamber:
1. With the inner and outer doors secure, press Start Cycle.
2. When the Sequence Complete light comes on, open the inner door
and transfer the material to the interchange.
3. Close and secure the inner door, then open the outer door to remove
any material.
4. Close and secure the outer door.
1-10
Anaerobic System
Thermo Scientific
Section 1
Installation and Start-Up
Automatic Operation
of Interchange (cont.)