Shelves
(2) adjustable, solid stainless steel in cabinet
18.8” W x 13.0” F-B (47.8cm x 32.0cm)
(2) solid stainless steel in incubator
17.5” W x 12.0” F-B (42.4cm x 30.5cm)
Vacuum Pump
Domestic . . . . . Stock # 901024 - 115VAC, 60Hz
Export . . . . . . . . Stock # 901025 - 220VAC, 50Hz
7.62 cfm, pulls down interchange to 20 in./Hg in 15 seconds. In-house
vacuum systems must meet 4 cfm, capacity to 24 in./Hg minimum
requirements.
Air Flow
(1) 27 CFM blower in the catalyst system
Internal Outlets
(4) outlets total, 1 foot-operated
Domestic
Load should not exceed 1.25 amps per outlet, or 5 amps total.
Export
Appliances should be 220V. Load should not exceed 0.5 amp per outlet,
or 2 amps total.
Electrical Requirements
Model 1025
90-130VAC, 1 Phase, 50/ 60Hz, 14.6 FLA @ 120V including vacuum
pump
Model 1029
180-235VAC, 1 Phase, 50/60Hz, 7 FLA @ 220V including vacuum
pump.
6-2
Anaerobic System
Thermo Scientific
Section 6
Specifications