1-2
Anaerobic System
Thermo Scientific
Section 1
Installation and Start-Up
Location
Locate the cabinet on a firm, level surface in an area of minimum
ambient temperature fluctuation. Allow adequate space at the rear and to
the side of the cabinet for connecting the gas supplies and the vacuum
pump. There should also be clearance behind the cabinet for the optional
foot switch. Penetration through the counter-top may be necessary to
accommodate the foot switch cord if the table is flush-mounted to the
wall.
Connect the interchange to an adequate house vacuum system or to an
intermittent vacuum pump. If the interchange is connected to a house
vacuum system, the system must meet the minimum requirements of 4
CFM capacities to 24 in/Hg.
It is recommended that the countertop house vacuum systems be
supplied with a 1/2” non-tapering nipple. Conventional tapered barb
nipples restrict vacuum volume.
Caution
If the system is to be connected to a house vacuum, refer to
Section 4 for instructions in reversing the direction of the vacuum
solenoid valve before connecting.
s
1. Select a location for the vacuum pump and secure the pump as
illustrated in Figure 1-2.
2. Install the nylon adapter (supplied) on the vacuum connection on the
rear plumbing panel.
Items Shipped Installed
Qty
Stock No.
Description
4
224249
12” Aluminum Bracket
2
92254245
Fluorescent Lamps
8
500174
Pilaster Clips, Stainless Steel
2
1025131
Chamber Shelf
2
1025274
Incubator Shelf
1
1025284
Incubator False Bottom
Connect the Vacuum
Source