![Tektronix DM 5010 Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/tektronix/dm-5010/dm-5010_instruction-manual_1078359047.webp)
Programming
—
DM
5010
DCV
(DC Voltage Function)
Type:
Setting
Setting
syntax:
DCV
< number
>
DCV
DIGIT
(Digital Resolution)
Type:
Setting
or
query
Setting syntax:
DIGIT
3.5
DIGIT 4.5
DIG
3.5
DIG
4.5
Examples:
DCV
1.5
DCV
DCV -1.E+3
Range selected:
2 V
1000
mV, auto-range
1000
V, auto-range
Discussion;
The header
selects the DCV function; the argument se
lects
a fixed voltage range. The format for numeric argu
ments
is
described under Number Format later
in this
section of
the
manual. The argument can be any value;
however,
the instrument rounds the argument up to the first
full
scale range. For instance, for an argument of 1.5, the
instrument
selects the
2 V
range.
If
the
argument is omitted or its value is 0 or less, the
instrument auto-ranges
beginning at the highest
range.
If the
argument is
above
the highest range, the instru
ment
generates a
command error (and asserts SRQ if RQS
is
on).
Ranges:
200
mV
2 V
20
V
200
V
1000
V
Query
syntax:
DIGIT?
DIG?
Query response:
DIGIT 3.5;
DIGIT 4.5;
Discussion:
This
command selects
the conversion rate. The argu
ment
3.5
sets the FAST conversion rate (3.5 digit resolu
tion). A
reading takes approximately 35 ms in the voltage
functions
and approximately 130 ms in the OHMS function.
The
argument 4.5 sets the normal conversion rate (4.5
digit
resolution). A reading takes approximately 310 ms in
the
voltage
functions;
620 ms
in the OHMS function.
REV
JAN
1982
DCV
DIGIT
3-9
Содержание DM 5010
Страница 14: ...DM 5010 2994 00 DM 5010 Programmable Digital Multimeter xii ADD JUL 1986...
Страница 27: ...Operating Instructions DM 5010 2994 03 Fig 2 3 DM 5010 front panel controls and connectors 2 3...
Страница 38: ......
Страница 40: ...Programming DM 5010 2994 07 Fig 3 1 instrument commands and relationship to front panel controls 3 2...
Страница 101: ...Theory of Operation DM 5010 SOURCE ACCEPTOR 2874 1 50A Fig 4 15 GP B Handshake flowchart ADD JAN 1982 4 27...
Страница 134: ......
Страница 208: ......
Страница 222: ......
Страница 250: ......
Страница 251: ...Section 8 DM 5010 OPTIONS No options are available 8 1...
Страница 252: ......
Страница 270: ......
Страница 272: ...DM 5010 2994 37 Fig 10 2 Location of DM 5010 adjustments and test points...
Страница 273: ......
Страница 274: ......
Страница 275: ......
Страница 276: ...DM 5010 2994 112 DM 5010 BLOCK DIAGRAM...
Страница 281: ......
Страница 282: ......
Страница 291: ......
Страница 293: ......
Страница 294: ......
Страница 297: ......
Страница 298: ......
Страница 303: ......
Страница 304: ...I...
Страница 305: ......
Страница 310: ......
Страница 311: ......
Страница 315: ......
Страница 318: ......
Страница 321: ......
Страница 323: ......
Страница 326: ......
Страница 332: ...2994 57...
Страница 334: ......
Страница 335: ......
Страница 336: ......
Страница 337: ...63 REV JUN 1986...
Страница 338: ...FIG 1 EXPLODED DM 5010...
Страница 339: ......
Страница 340: ......
Страница 341: ......
Страница 347: ......