Specification
—DM 5010
Performance
Check
in this manual. Information given
in the
Supplemental Information
and Description columns of
the
following
tables is provided for user information only and
should
not
be interpreted to be Performance Check require
ments.
The information
under Electrical Characteristics ap
plies to
both front-panel and rear-interface measurements,
unless otherwise noted.
NOTE
For
measurements to
specified accuracy, internal ad
justments should be performed after
1000 hours
of
operation or every six
months if used infrequently. Re
fer
to the Adjustment
Procedure in this manual.
Table
1-2
ELECTRICAL CHARACTERISTICS
Characteristics
Performance
Requirements
Supplemental
Information
DC VOLTS
ACCURACY for
the 200 mV,
2 V,
20
V, 200
V, and 1000 V ranges:
Normal
Conversion Rate
(3
readings per second)
+
18°C to
+28°C
200 mV
range
± (0.015% of reading
+ 0.01% of full
scale)
2
V
through
200 V ranges
±(0.015% of
r
0.005% of full
scale)
1000 V range
±(0.020% of reading
+ 0.010% of full
scale)
0°C to
+18°C,
+ 28°C
to
+ 50°C
200
mV
range
±(0.06%
of
r 0.035% of full
scale)
For the typical accuracy for a given tem
perature
in this range, add to the
+18°C
to
-r28°C
accuracy, ±(0.002% of read
ing + 0.001% of full scale)/0C
deviation
from
+28°C or +1B°C.
2 V through 200 V ranges
±(0.06%
of r 0.03% of full
scale)
1000 V range
±(0.065% of
r
0.035% of full
scale)
Fast
Conversion
Rate
(26
readings per second)
+
18°C to
+28°C
200
mV to 200 V
ranges
±(0.05% of reading
+ 0.05% of full
scale)
1000 V range
±(0.05% of reading
+ 0.1% of full scale)
0°C to
+ 18°C,
+ 28°C to
+ 50°C
200 mV to 200 V ranges
±(0.1% of
reading
+ 0.1% of full scale)
1000 V
range
±(0.1% of
reading
+ 0.15% of full scale)
1-2
Содержание DM 5010
Страница 14: ...DM 5010 2994 00 DM 5010 Programmable Digital Multimeter xii ADD JUL 1986...
Страница 27: ...Operating Instructions DM 5010 2994 03 Fig 2 3 DM 5010 front panel controls and connectors 2 3...
Страница 38: ......
Страница 40: ...Programming DM 5010 2994 07 Fig 3 1 instrument commands and relationship to front panel controls 3 2...
Страница 101: ...Theory of Operation DM 5010 SOURCE ACCEPTOR 2874 1 50A Fig 4 15 GP B Handshake flowchart ADD JAN 1982 4 27...
Страница 134: ......
Страница 208: ......
Страница 222: ......
Страница 250: ......
Страница 251: ...Section 8 DM 5010 OPTIONS No options are available 8 1...
Страница 252: ......
Страница 270: ......
Страница 272: ...DM 5010 2994 37 Fig 10 2 Location of DM 5010 adjustments and test points...
Страница 273: ......
Страница 274: ......
Страница 275: ......
Страница 276: ...DM 5010 2994 112 DM 5010 BLOCK DIAGRAM...
Страница 281: ......
Страница 282: ......
Страница 291: ......
Страница 293: ......
Страница 294: ......
Страница 297: ......
Страница 298: ......
Страница 303: ......
Страница 304: ...I...
Страница 305: ......
Страница 310: ......
Страница 311: ......
Страница 315: ......
Страница 318: ......
Страница 321: ......
Страница 323: ......
Страница 326: ......
Страница 332: ...2994 57...
Страница 334: ......
Страница 335: ......
Страница 336: ......
Страница 337: ...63 REV JUN 1986...
Страница 338: ...FIG 1 EXPLODED DM 5010...
Страница 339: ......
Страница 340: ......
Страница 341: ......
Страница 347: ......