Operating Instructions—DM 5010
Fig. 2-6, Setup for calculation example 2.
Set
the
sinewave generator to mid-range (5 kHz
in this
example); adjust the sinewave generator amplitude for a 1 V
reading
on the DM
5010.
Press
the DM 5010 dBr
button. The display reads 0.0.
Reduce the
sinewave generator frequency until the
DM
5010 display
reads
—3.00. (Do not readjust
amplitude.)
The frequency
of the generator is the lower
-3 dB point of
the audio amplifier.
Example
3:
Using
COMPARE to select resistors within
2%
of
the nominal value.
The
COMPARE and X-B/A calculations may be com
bined
in the above example. This combination eliminates fig
uring
the
highest and lowest in-toierance values; only the
nominal resistance value and the tolerance are
used as
constants.
Set
constant B to 15000 (nominal
resistance).
Set
constant
A to 150 where A = B (.01). This converts
the
difference between nominal value
and actual value to a
per cent.
Set
one LIMITS
constant to
2 (for a 2% tolerance).
Set the DM 5010
front-panel
controls as follows:
Set
the
other
LIMITS
constant to
—2.
OHMS
on
NULL
off
LOW
FREQ
RESPONSE
off
RANGE
AUTO
TRIGGER MODE
RUN
CONVERSION RATE
FAST off
CALCULATIONS
COMPARE
on
all others
off
REAR
INPUT
off
To select 15 kfi resistors within 2% of the nominal value,
set one
LIMITS constant to 15300. Set the other LIMITS
constant
to 14700. Connect the first resistor
to the
DM
5010 front-panel HIGH
and LOW input connectors.The
DM
5010 displays HI or
LO
if the resistor is
above or below
the
2% tolerance. PASS is displayed if the resistor is be
tween
or equal
to the limits.
Press
X-B/A.
The DM 5010 displays PASS, Hi, or LO.
Repackaging Information
If
this
Tektronix instrument is to be shipped to a
Tektronix
Service Center for service or repair, attach a tag
showing
owner
(with
address) and the name of
an individual
at
your firm that can be contacted. Include the complete
instrument serial number
and a description of the service
required.
Save
and re-use
the package in which the instrument
was
shipped.
If the original packaging is unfit for use or not
available, repackage the instrument
as follows:
2-12
Содержание DM 5010
Страница 14: ...DM 5010 2994 00 DM 5010 Programmable Digital Multimeter xii ADD JUL 1986...
Страница 27: ...Operating Instructions DM 5010 2994 03 Fig 2 3 DM 5010 front panel controls and connectors 2 3...
Страница 38: ......
Страница 40: ...Programming DM 5010 2994 07 Fig 3 1 instrument commands and relationship to front panel controls 3 2...
Страница 101: ...Theory of Operation DM 5010 SOURCE ACCEPTOR 2874 1 50A Fig 4 15 GP B Handshake flowchart ADD JAN 1982 4 27...
Страница 134: ......
Страница 208: ......
Страница 222: ......
Страница 250: ......
Страница 251: ...Section 8 DM 5010 OPTIONS No options are available 8 1...
Страница 252: ......
Страница 270: ......
Страница 272: ...DM 5010 2994 37 Fig 10 2 Location of DM 5010 adjustments and test points...
Страница 273: ......
Страница 274: ......
Страница 275: ......
Страница 276: ...DM 5010 2994 112 DM 5010 BLOCK DIAGRAM...
Страница 281: ......
Страница 282: ......
Страница 291: ......
Страница 293: ......
Страница 294: ......
Страница 297: ......
Страница 298: ......
Страница 303: ......
Страница 304: ...I...
Страница 305: ......
Страница 310: ......
Страница 311: ......
Страница 315: ......
Страница 318: ......
Страница 321: ......
Страница 323: ......
Страница 326: ......
Страница 332: ...2994 57...
Страница 334: ......
Страница 335: ......
Страница 336: ......
Страница 337: ...63 REV JUN 1986...
Страница 338: ...FIG 1 EXPLODED DM 5010...
Страница 339: ......
Страница 340: ......
Страница 341: ......
Страница 347: ......