DM
5010 Instrument
Interfacing
Guide
Table
2
(cont)
Header
Argument
Description
ERR?
Returns
error
code for most
recent
event
reported by seri
al poll when RQS is ON; with
RQS
OFF it returns the high
est
priority status.
FUNCT?
Returns the
current
function
(FUNC?)
(DCV, OHMS,
DIODE, ACV,
ACDC)
and range. Negative
argument indicates auto
range.
ID?
Returns
"ID
TEK/
DM5010,V79.1
FXX;’, where
XX is the
firmware
version
number.
INIT
Initializes
all instrument
settings
to
their power-on
settings. See Power On
Settings.
LFR
ON
Enables
the
LOW FREQ RE
SPONSE
function.
Instrument
computes the average of four
ACV
or
ACV+DCV
measurements.
LFR
OFF
Disables the LOW FREQ RE
SPONSE function.
LFR?
Returns
“
LFR OFF;
’ or “LFR
ON;
’.
LIMITS
<num>,
Sets
limits used
in CALC
(LIM)
<num>
CMPR
program.
LIMITS?
Returns
"LIMITS
<num>,
(LIM?)
<num>;
’.
MODE
RUN
Selects
the
free-run
(RUN)
trigger
mode.
MODE
TRIG
Selects TRIGGERED
trigger
mode.
Triggers a
conversion
only
upon
receipt of one of
the following.
•
SEND
command
•
Group Execute Trigger
<GET>, only if device
trigger
(DT)
is enabled.
•
My
Talk Address (MTA)
with
the
output
unspecified
(no query command).
Table
2
(cont)
Header
Argument
Description
•
Rear interface trigger
(EXTRIG). Requires inter
nal jumper installation. To
cause a single trigger, this
line
must remain low be
tween
0.5
and 10 µsec.
Holding
this
line low longer
causes
multiple readings.
MODE?
Returns
“MODE RUN;" or
“MODE
TRIG;’.
MONITOR
ON
Enables
monitor SRQ. Saves
(MON)
the measurement outside the
limits set by the LIMITS com
mand and generates
an SRQ.
Returns
this measurement
in
response
to
DATA. Subse
quent
out-of-limits measure
ments
are not
reported until
the
SRQ is serviced and
mea
surement is
returned via
DATA
command.
MONITOR
(MON)
OFF
Disables
the monitor
SRQ.
MONITOR?
Returns
"MONITOR ON;" or
(MON?)
“MONITOR OFF;'.
NULL
<num>
Sets
an offset value and ap
plies it to subsequent
mea
surements. Argument may
be
any value
up to 100%
of
range.
Disable
NULL
by set
ting argument to
0, or chang
ing measurement function.
NULL?
Returns
“NULL
<num>;".
OHMS
<num>
Selects
OHMS function.
Ar
gument
selects range.
Nega
tive or omitted
argument
selects auto-range.
OPC
ON
Enables
operation
complete
SRQ.
DM
5010
asserts SRQ
whenever
a new measure
ment is available.
OPC
OFF
Disables
operation complete
SRQ.
OPC?
Returns
“
OPC OFF;’ or
"OPC ON;’
.
4
Содержание DM 5010
Страница 14: ...DM 5010 2994 00 DM 5010 Programmable Digital Multimeter xii ADD JUL 1986...
Страница 27: ...Operating Instructions DM 5010 2994 03 Fig 2 3 DM 5010 front panel controls and connectors 2 3...
Страница 38: ......
Страница 40: ...Programming DM 5010 2994 07 Fig 3 1 instrument commands and relationship to front panel controls 3 2...
Страница 101: ...Theory of Operation DM 5010 SOURCE ACCEPTOR 2874 1 50A Fig 4 15 GP B Handshake flowchart ADD JAN 1982 4 27...
Страница 134: ......
Страница 208: ......
Страница 222: ......
Страница 250: ......
Страница 251: ...Section 8 DM 5010 OPTIONS No options are available 8 1...
Страница 252: ......
Страница 270: ......
Страница 272: ...DM 5010 2994 37 Fig 10 2 Location of DM 5010 adjustments and test points...
Страница 273: ......
Страница 274: ......
Страница 275: ......
Страница 276: ...DM 5010 2994 112 DM 5010 BLOCK DIAGRAM...
Страница 281: ......
Страница 282: ......
Страница 291: ......
Страница 293: ......
Страница 294: ......
Страница 297: ......
Страница 298: ......
Страница 303: ......
Страница 304: ...I...
Страница 305: ......
Страница 310: ......
Страница 311: ......
Страница 315: ......
Страница 318: ......
Страница 321: ......
Страница 323: ......
Страница 326: ......
Страница 332: ...2994 57...
Страница 334: ......
Страница 335: ......
Страница 336: ......
Страница 337: ...63 REV JUN 1986...
Страница 338: ...FIG 1 EXPLODED DM 5010...
Страница 339: ......
Страница 340: ......
Страница 341: ......
Страница 347: ......