Programming
—DM 5010
COMMANDS
The
instrument is controlled by the
front panel or
via
commands
received from the controller.
These commands
are
of three
types:
Setting commands—
control
instrument settings.
Query-output commands
—
ask for
data.
Operational commands—
cause
a
particular action.
The
instrument responds to
and executes all commands
when
in the remote state.
When in the local state,
setting
and
operational
commands
generate errors
since instru
ment
functions are under front panel control; only
query
output commands
are
executed.
Each command
begins with a header
—a word
that de
scribes
the
function
implemented.
Many commands require
an argument following the header—
a
word or number that
specifies the desired state for the function.
Using
fewer characters than the abbreviated header
or
argument
should be done with caution since erro
neous
results or damage could result if this data is
sent to the wrong instrument.
>
CAUTION
<
With
the instrument in the AUTO range mode, do not
repeatedly
switch the input voltage between a
low val
ue (<200 mV
peak) and a higher value (>200 V
peak).
For
repeated measurements
alternating be
tween
these voltage extremes, use the STEP range
mode to select an
appropriate higher range before in
creasing
the input voltage.
Failure to do so
may tem
porarily
cause inaccuracy in measurements made
using
the 200 mV
range.
FUNCTIONAL
COMMAND LIST
INSTRUMENT
COMMANDS
Function Commands
ACDC
<num>
—Sets
the
ACV + DCV function and
range.
ACV
<num>—Sets the ACV function and range.
DCV
<num>—Sets
the DCV function and
range.
DIODE
—Sets the DIODE TEST function.
FUNCT?—Returns present function and
range.
LFR
ON—Enables
the LOW FREQ RESPONSE
function.
LFR
OFF—Disables the LOW FREQ RESPONSE
function.
LFR?—
Returns
LFR
ON or LFR
OFF.
NULL
<num>—Enables the NULL function and offset
value.
NULL?
—Returns NULL offset value.
OHMS
<num>—Sets the OHMS function and range.
Trigger
Mode
Commands
MODE
RUN—Enables
the
RUN trigger mode.
MODE TRIG—Enables the TRIGGERED trigger
mode.
MODE?—Returns
MODE
RUN or MODE TRIG.
RDY?—Returns RDY 1
if a measurement is ready; RDY
0 if one is
in progress or waiting for trigger.
DIGIT 3.5—Enables
FAST
conversion rate.
DIGIT 4.5—Enables normal conversion rate.
DIGIT?—Returns DIGIT 3.5 or DIGIT 4.5.
Calculation Commands
AVE
<num>—Sets the value of constant N.
AVE?—Returns value of constant N.
CALC AVE
—Enables the AVERAGE calculation.
CALC CMPR—Enables the COMPARE calculation.
CALC
DBM—Enables the dBm calculation.
CALC DBR—Enables
the dBr calculation
3-3
CAUTION
Содержание DM 5010
Страница 14: ...DM 5010 2994 00 DM 5010 Programmable Digital Multimeter xii ADD JUL 1986...
Страница 27: ...Operating Instructions DM 5010 2994 03 Fig 2 3 DM 5010 front panel controls and connectors 2 3...
Страница 38: ......
Страница 40: ...Programming DM 5010 2994 07 Fig 3 1 instrument commands and relationship to front panel controls 3 2...
Страница 101: ...Theory of Operation DM 5010 SOURCE ACCEPTOR 2874 1 50A Fig 4 15 GP B Handshake flowchart ADD JAN 1982 4 27...
Страница 134: ......
Страница 208: ......
Страница 222: ......
Страница 250: ......
Страница 251: ...Section 8 DM 5010 OPTIONS No options are available 8 1...
Страница 252: ......
Страница 270: ......
Страница 272: ...DM 5010 2994 37 Fig 10 2 Location of DM 5010 adjustments and test points...
Страница 273: ......
Страница 274: ......
Страница 275: ......
Страница 276: ...DM 5010 2994 112 DM 5010 BLOCK DIAGRAM...
Страница 281: ......
Страница 282: ......
Страница 291: ......
Страница 293: ......
Страница 294: ......
Страница 297: ......
Страница 298: ......
Страница 303: ......
Страница 304: ...I...
Страница 305: ......
Страница 310: ......
Страница 311: ......
Страница 315: ......
Страница 318: ......
Страница 321: ......
Страница 323: ......
Страница 326: ......
Страница 332: ...2994 57...
Страница 334: ......
Страница 335: ......
Страница 336: ......
Страница 337: ...63 REV JUN 1986...
Страница 338: ...FIG 1 EXPLODED DM 5010...
Страница 339: ......
Страница 340: ......
Страница 341: ......
Страница 347: ......