![Tektronix DM 5010 Скачать руководство пользователя страница 42](http://html1.mh-extra.com/html/tektronix/dm-5010/dm-5010_instruction-manual_1078359042.webp)
Programming—DM
5010
CALC
RATIO—
Enables X-B/A calculation.
CALC
OFF—Disables all calculations.
CALC?
—Returns CALC OFF or the enabled
calculation(s).
DBR <num>—Sets the value of the ref constant.
DBR?—Returns value of ref constant.
LIMITS
<num>,<num>
—Sets values of LIMITS
constants.
LIMITS?—
Returns values of
LIMITS constants.
MONITOR ON—
Enables
SRQ when
measurement
exceeds
LIMITS
constants.
MONITOR
OFF
—Disables SRQ when
measurement
exceeds
LIMITS constants.
MONITOR?—Returns
MONITOR ON or
MONITOR
OFF.
RATIO
<num>,<num>—Sets values of A and B
constants.
RATIO?—Returns
values of A and B constants.
INPUT/OUTPUT
COMMANDS
DATA
—Outputs data saved by MONITOR SRQ.
SEND
—Outputs
data in
Output Buffer; triggers, if
necessary.
SOURCE
REAR
—Selects rear interface connector input.
SOURCE
FRONT—Selects
front panel connector input.
SOURCE?—Returns
SOURCE FRONT
or SOURCE
REAR.
SYSTEM COMMANDS
DT TRIG—Enables device
trigger function. Instrument
triggers
after
<GET>
interface message.
DT OFF—
Disables
device trigger function.
DT?—Returns
DT
TRIG or DT
OFF.
ERR?—Returns error code.
ID?—Returns instrument
identification and firmware
version.
INIT—Initializes
instrument settings.
SET?—Returns
instrument settings.
TEST—Returns 0 for correct calibration
checksum;
351
for incorrect.
STATUS
COMMANDS
OPC ON—Enables operation complete SRQ.
OPC OFF—Disbles operation complete SRQ.
OPC?—
Returns
OPC ON or OPC
OFF.
OVER
ON—Enables overrange SRQ.
OVER OFF
—Disables overrange SRQ.
OVER?—Returns OVER
ON or OVER OFF.
RQS ON—
Enables
generation of SRQ s.
RQS OFF—Disables
generation of SRQ’s.
RQS?—Returns
RQS ON or
RQS OFF.
USER ON
—Enables SRQ when ID
button is pushed.
USER
OFF—Disables SRQ when ID button is pushed.
USER?—Returns
USER ON or USER
OFF.
3-4
Содержание DM 5010
Страница 14: ...DM 5010 2994 00 DM 5010 Programmable Digital Multimeter xii ADD JUL 1986...
Страница 27: ...Operating Instructions DM 5010 2994 03 Fig 2 3 DM 5010 front panel controls and connectors 2 3...
Страница 38: ......
Страница 40: ...Programming DM 5010 2994 07 Fig 3 1 instrument commands and relationship to front panel controls 3 2...
Страница 101: ...Theory of Operation DM 5010 SOURCE ACCEPTOR 2874 1 50A Fig 4 15 GP B Handshake flowchart ADD JAN 1982 4 27...
Страница 134: ......
Страница 208: ......
Страница 222: ......
Страница 250: ......
Страница 251: ...Section 8 DM 5010 OPTIONS No options are available 8 1...
Страница 252: ......
Страница 270: ......
Страница 272: ...DM 5010 2994 37 Fig 10 2 Location of DM 5010 adjustments and test points...
Страница 273: ......
Страница 274: ......
Страница 275: ......
Страница 276: ...DM 5010 2994 112 DM 5010 BLOCK DIAGRAM...
Страница 281: ......
Страница 282: ......
Страница 291: ......
Страница 293: ......
Страница 294: ......
Страница 297: ......
Страница 298: ......
Страница 303: ......
Страница 304: ...I...
Страница 305: ......
Страница 310: ......
Страница 311: ......
Страница 315: ......
Страница 318: ......
Страница 321: ......
Страница 323: ......
Страница 326: ......
Страница 332: ...2994 57...
Страница 334: ......
Страница 335: ......
Страница 336: ......
Страница 337: ...63 REV JUN 1986...
Страница 338: ...FIG 1 EXPLODED DM 5010...
Страница 339: ......
Страница 340: ......
Страница 341: ......
Страница 347: ......