Amplifiers
4-25
■
Settings
The following dialog box appears when you choose
Setup
and
Params
, and
then double-click the channel that you want to set.
●
Zero Balancing
Click
Exec
to perform zero balancing whenever
the GX-1 is powered on.
After selecting the option
All Strain Amp
and
clicking
Exec
, zero balancing is performed
simultaneously on all channels on the installed
dynamic strain input amps.
▼
Disable post trigger before zero balancing.
▼
Enable both channels of a slot.
▼
Filter cutoff frequency is set slot by slot, not channel
by channel.
Channel
:
Select IN to record with this channel, or select DISABLE not
to record with this channel.
Range
:
Select the input range.
Sample Frequency
: You can select 10 times the
Base Sample Freq
when you turn on
Multi Sampling
check box.
Channel Name
: You can name the channel. Use up to 16 alphanumeric
characters.
Filter
:
Select the filter cutoff frequency.
Bridge Voltage
: Select 2 V or 5 V for the excitation voltage.
All Strain Amp set same condition
: When this option is selected, the same
parameters are set for all signal conditioners of the same type.
Содержание GX-1
Страница 5: ...iv...
Страница 6: ...Introduction 1 1 Overview 1 3 Features 1 4 About TAFFmat 1 6 Precautions 1 6 Name of Each Part 1 7...
Страница 7: ...1 2...
Страница 15: ...1 10...
Страница 17: ...2 2...
Страница 27: ...2 12...
Страница 28: ...GX Navi 3 1...
Страница 29: ...3 2...
Страница 47: ...3 20...
Страница 67: ...3 40...
Страница 68: ...Technical Information 4 1...
Страница 69: ...4 2...
Страница 79: ...Specifications 4 12 External Dimensions Unit mm 300 85 200...
Страница 80: ...4 13...
Страница 81: ...4 14...
Страница 116: ...Appendix 5 1 In Case of Trouble 5 3 Index 5 4 Supplied Accessories 5 5 Separately Purchasable Options 5 5...
Страница 117: ...5 2...
Страница 121: ...5 6...