69
1 5 . 2
–
R
E M O T E
A
L A R M A N D
A
U T O
-
D I A L E R
An automated dialer can be attached which will automatically dial a pre-set number if
an alarm occurs during a run.
The socket of the Remote Alarm connector has 3 contacts with the following
disposition:
No alarm
= contacts 1-2 closed, contacts 1-3 open.
Alarm in course
= contacts 1-2 open, contacts 1-3 closed.
Because there are no voltages of any kind present on these contacts (they are
isolated from the rest of the instrument), they can be utilized to activate an external
device such as an Auto-Dialer to call a pre-selected number and relay the message
that an alarm has occurred.
If you wish to attach such a device to the ATP™ but do not wish to purchase our
Auto-Dialer, or, if you have a proprietary system already in place, please note the
following constraints on the connector and its associated electronics:
Maximum allowed voltage: 48V AC/DC
Maximum allowed current: 6A
For instructions related to the Auto-Dialer, please consult the manual included with
the dialer.