55
1 2 - F I L L T I M E O U T
The filling from a reagent or paraffin bottle has not occurred in the allotted time.
Cause –
See Alarm 11.
Solution –
See Alarm 11.
1 3 - O V E R F I L L S
An overfill of reagent or paraffin has occurred in the Retort (detected by the overfill
sensor).
Cause –
Incomplete emptying of previous reagent due to a solenoid valve
failure or a paraffin temperature problem.
Solution –
No intervention by the user is possible. If the problem persists,
contact the TBS
®
service department for technical assistance.
1 4 - O V E R T
°
R E T O R T
During a paraffin step the Retort temperature went over the allowed limit (not to be
confused with alarm #20).
Cause –
Thermostat system damage.
Solution –
No intervention by the user is possible. If the problem persists,
contact the TBS
®
service department for technical assistance.
1 5 - T S C O U T
Malfunction of the retort temperature sensor.
Cause –
TSC sensor failure.
Solution –
No intervention by the user is possible. If the problem persists,
contact the TBS
®
service department for technical assistance.
1 6 - S H O R T F I L L
A fill is completed before the minimum allotted time. The retort will drain and the
process will proceed to the next step. If the problem persists, the process will be
stopped.
Cause –
The reagent bottle quick connector is not fully inserted into the
female receptacle; inefficient air pump.
Solution –
Check reagent bottle position and fluid levels. If it happens on the
alcohol stage of the Purge Cycle, reduce the percentage of the alcohol to
80%. If the problem persists, contact the TBS
®
service department for
technical assistance
1 7 - V R O U T
The rotary valve is out of position.