58
Cause –
See Alarm 30.
Solution –
See Alarm 30.
3 2 - C O M E R R O R [ O V E R R U N ]
See Alarm 30.
Cause –
See Alarm 30.
Solution –
See Alarm 30.
3 3 - C O M E R R O R [ T I M E O U T ]
See Alarm 30.
Cause –
See Alarm 30.
Solution –
See Alarm 30.
3 9 - C O M E R R O R [ L I N E O R P O R T D O W N ]
See Alarm 30.
Cause –
See Alarm 30.
Solution –
See Alarm 30.
4 0 - E L O D E V L A C K
A failure of the touch screen. The display will still be functional.
Cause –
A failure has occurred in the Touch Screen membrane circuit, power
supply, connectors, CPU board, or Disk-on-Chip.
Solution –
A standard computer keyboard may be connected to the 5 pin din
connector keyboard connector next to the remote alarm connector on the
upper rear of the ATP™. All the functions of the ATP™ can be operated from
the keyboard if the display screen is working, but not the touch screen
membrane. Contact the TBS
®
service department for assistance.
5 0 - O V E R P R E S S U R E
The purpose of this alarm is to exclude the clogging of a pneumatic line, which
greatly reduces the volume of the overall plumbing system, from the causes of
alarms # 2, 4, 5, 8, 11, 12, 23, 24, 25, 27.
Cause –
A blockage in a pneumatic system of the ATP™ prior to the Retort.
The problem may be a bad connection or a faulty signal from the pressure
transducer. The pressure transducer may have failed.
Solution –
Check: the pneumatic plumbing system between the pump and the
Retort for blockages. Contact the TBS
®
service department for assistance.