56
Cause –
The Rotary Valve motor has failed or the Rotary Valve position
sensor has failed.
Solution –
No intervention by the user is possible. If the problem persists,
contact the TBS
®
service department for technical assistance
1 8 - F I L E N O T F O U N D
An essential computer file is damaged or cannot be found and the error recovery
system failed trying to correct the problem.
Cause –
Computer malfunction, back-up battery failure.
Solution –
No intervention by the user is possible. If the problem persists,
contact the TBS
®
service department for technical assistance
1 9 - R E T O R T T ° L O W
The retort temperature was too low during the drain following a paraffin step, or the
drain cannot be completed. The process will be stopped to avoid other problems.
(See also Alarm 2.)
Cause –
Retort heating system failure.
Solution –
No intervention by the user is possible. If the problem persists,
contact the TBS
®
service department for technical assistance
2 0 - W A X T ° L O W
During the fill of a paraffin, the temperature of the retort was too low. The process
will be halted because the fill cannot be guaranteed.
Cause –
See Alarm 14.
Solution –
See Alarm 14.
2 1 - O V E R T ° W A X
The temperature of the paraffin chamber has exceeded the set point.
Cause –
Temperature system failure.
Solution –
No intervention by the user is possible. If the problem persists,
contact the TBS
®
service department for technical assistance
2 2 - T S W O U T
The electrical signal from the CCW temperature sensor is out of the correct range.
Cause –
The sensor has malfunctioned.
Solution –
No intervention by the user is possible. If the problem persists,
contact the TBS
®
service department for technical assistance
2 3 - P R E S S U R E N . R .
During the first phase of a drain, the pressure in the retort has not reached the
correct value. The process will be stopped.
Содержание ATP1-120
Страница 80: ...73 APPENDIX 4 ATP PROGRAM MENU FLOW...