57
Cause –
The air pump or other pneumatic circuit parts have malfunctioned.
Solution –
No intervention by the user is possible. If the problem persists,
contact the TBS
®
service department for technical assistance.
2 4 - D R A I N N O P R E S S
During the second phase of a drain, the pressure in the retort has not reached the
correct value. The process will be stopped.
Cause –
See Alarm 23.
Solution –
See Alarm 23.
2 5 - D R A I N T I M E O U T
The drain has gone over the maximum time allowed. The process will be stopped.
Cause –
See Alarm 23.
Solution –
See Alarm 23.
2 6 - R E T . N O T E M P T Y
The user has attempted to begin a process with the retort either not empty or without
having run a Purge Cycle following the previous process.
Cause –
Due to another alarm or due to an interrupted process, the retort is
not empty. Also caused by neglecting to purge after a process.
Solution –
Clean the retort, if it is not empty, drain it manually. At the
Service Menu perform the “
Alarm Reset
” and the “
Flag Reset
”. Return to the
Main Menu and start the Purge Cycle using the “CLEAN” key.
2 7 - O V E R F I L L P
An overfill of reagent or paraffin has occurred in the Retort. (Alarm sensed by fast
pressure gradient during fill.)
Cause –
Complete emptying of the previous reagent failed.
Solution –
No intervention by the user is possible. If the problem persists,
contact our technical assistance
3 0 - C O M E R R O R [ F R A M E ]
A communication error between the CPU and serial board.
Cause –
A component or connection in the computer has failed
.
Solution –
Contact the TBS
®
service department for assistance.
3 1 - C O M E R R O R [ P A R I T Y ]
See Alarm 30.