The
End-user Service Menu:
V a c u u m t e s t
The instrument will create a vacuum as if a program is running; this test should be
done with the retort empty.
•
The instrument should reach a pressure between 300 and 350mBar (TP field)
within 50 to 60 seconds (GRADIENT field).
If the GRADIENT value exceeds 60 seconds, the Retort lid gasket should be
checked and cleaned. If this does not correct the problem, contact the TBS
®
service
department for technical service.