When defining the process number/cassette number limits for groups of reagents,
the same factors that were considered when assigning single bottle limits will be
considered here. The difference is that when the preset limit for the group is
reached or exceeded, the
RMS
will require the substitution of the bottle whose
reagent is the most polluted (the bottle in the group that, in the previous processes,
was used first).
9 . 5
–
R M S
–
C
O U N T E R S
T
O T A L
R
E S E T
The Counters Total Reset selection allows the user to set to zero the number of
processes performed and the number of cassettes processed. This function may be
used when it is determined that it is necessary to replace/renew the reagents of all
bottles and, in effect, restart the
RMS
. The complete zero resetting
does not
include
the counters for the charcoal filters or the bottles of the Purge reagents.
Please note:
Resetting the counters to zero occurs automatically
when the following functions are selected:
•
Changing the
RMS
mode (between Single and Group mode)
•
Each time the
RMS
is activated or deactivated
•
When the definition and/or limits of the groups are modified
Please note:
Resetting the counters to zero does
not
occur
automatically when modifications to the definition of the reagents or of
the limits in Single Bottle mode are made.
R M S – S i n g l e C o u n t e r R e s e t
The Single Counter Reset selection allows the user to zero the counters of individual
bottles. This function is useful if the user decides to replace/renew a reagent before
the preset limit of a particular bottle is reached.
Please note:
The Single Counter Reset only allows the user to
change individual bottle counters if the
RMS
is in Single Bottle mode.
Please note:
The Single Counter Reset is also the only function that
allows the operator to reset the counters of the charcoal filters and the
Purge reagents.
Please note:
The counters of the charcoal filters and Purge reagents
are
not
reset by the Counters Total Reset function or a change from
Single Bottle mode to Group mode. Whenever the user replaces the
Purge reagents or the charcoal filters without being prompted by the
RMS,
it will be necessary to utilize the Single Counter Reset.