60
1 1 . 5
–
I
N S T R U M E N T
R
E S E T
In the event the software has locked up and the instrument ceases operation, the
software can be reset by the following procedure:
On the rear of the instrument there is a green reset button. Its function is to reset
the software under the following conditions:
•
If, for any reason, the user is unable to reset an alarm from the keyboard
•
If a user faces some other unexpected situation that cannot be resolved by
any other means
•
To reset that part of the instrument’s memory that holds variables pertaining
to the state and content of the retort, possible steps in process, possible
actions in process and/or alarms in process
The procedure for using the reset button is as follows:
•
Switch off the instrument (the switch is located where the power cord enters
the tissue processor).
•
Press and hold
the green button in the down position.
•
Switch on the instrument,
maintaining the green button in the down
position
until such time as the Main Menu appears on the LCD screen again.
After this operation, the ATP™ will be ready for operation again, but first the user
MUST
verify the following:
ü
That the retort is empty.
ü
That the retort is clean (above all in reference to the paraffin).
ü
That the lid gasket of the retort is clean and in its correct position.
ü
That the instrument works correctly, by performing tests from the service
menu.
ü
That an alarm present before the reset is not again displayed (in this case will
be necessary to switch off the instrument and call the TBS
®
service
department, with the exception of an alarm #1).
ü
It is advisable to run a Purge Cycle and/or an abbreviated process before
attempting a sample run in order to verify the correct operation of the
processor.
ü
Be certain to perform the checklist (located in Chapter 10.8) before starting a
program
or
a Purge Cycle.
Содержание ATP1-120