![tbs electronics ATP1-120 Скачать руководство пользователя страница 32](http://html1.mh-extra.com/html/tbs-electronics/atp1-120/atp1-120_instruction-manual_3820869032.webp)
25
If you are operating the ATP™ in an area with high relative humidity, if you run long
processes frequently, if you apply pressure and/or vacuum to most or all of the
process steps, if you agitate the reagents in the retort frequently, and especially if
several of these factors apply to your usage, it will be necessary to set your process
number limits for the charcoal filters at or below the minimum values mentioned
above.
Important note:
If you replace your charcoal filters before the
process counter reaches your preset limit, you must reset the counter
with the “RMS Single Counter Reset” function on the RMS Setup
menu. If this becomes a common practice, please edit you process
limits for the charcoal filters.
7 . 1 0
–
W P C
( W
A X
P
U R I F I C A T I O N
C
Y C L E
)
A
C T I V A T I O N
/
D
E A C T I V A T I O N
Use of the
WPC
will reduce your consumption of paraffin by approximately 50%.
The system is able to reduce paraffin consumption by removing the preceding
reagents that would be introduced by cross-contamination. This removal is
accomplished by a flow of air bubbled through the paraffin containers. The volatile
reagents evaporate due to the heating of the paraffin containers during the
WPC
process. The air bubbled through the paraffin carries these reagent vapors through
the charcoal filters where they are trapped. Use of the Paraffin Wax Purification
Cycle, while it will reduce consumption of paraffin waxes, will also shorten the
estimated life of the charcoal filters.
The duration of the Paraffin Wax Purification Cycle is about 30 minutes.
If the
WPC
option is activated, it will begin automatically after each
Purge cycle, but only when paraffin has been used in the preceding
process.
It is always possible to manually run a Paraffin Wax Purification Cycle from the
RMS
Setup Menu, independent of paraffin usage in the previous process. During the
WPC
the paraffin waxes are not transferred from their containers; the retort will stay
clean and it will not be necessary to start a Purge cycle after the
WPC
.
During the
WPC
, the retort lid must
always stay closed
.
The activation / deactivation of the
WPC
takes place on the RMS Setup menu. The
WPC
should be activated for optimal performance.