16
CHAPTER
6
–
PREPARATION
FOR
OPERATION
6 . 1
–
F
I L L I N G
R
E A G E N T
B
O T T L E S
The emptying and subsequent refilling of reusable reagent
bottles must be done in accordance with all safety
regulations for handling flammable and toxic substances.
Use proper ventilation and keep away from open flames
and/or electrical circuits.
1. Place bottles on a flat level surface for filling. All Bottles
must
be filled to the
level indicated by the markings on the bottle (between the Min and Max
marks):
MIN = 2.5 liters, MAX = 2.75 liters.
Failure to fill to the correct level can result in errors and alarms.
2. Inspect the Quick connect O-rings and inspect the internal tubing connection
to ensure there are no cracks or defects, and that the O-rings are clean.
3. Inspect the openings in the Quick connect to make sure there are no clogs or
obstructions.
When exchanging or refilling bottles, use the special screw-caps
supplied with the instrument to avoid reagent spillage and fumes.
Replace cracked or damaged caps and O-rings.
4. Secure the cap around the Quick connect, and the solid cap.
DO NOT OVER-TIGHTEN. This can result in cracking of the caps.
If the quick connect couplers leak, the o-rings should be replaced. If
leakage problems persist, contact TBS
®
customer service.