If there has been a power loss or the unit has been powered off and on, the alarm
reset will need to be used to clear the Blackout Alarm.
1 2 . 2
–
S
E R V I C E
M
E N U
A
B B R E V I A T I O N
L
I S T
On the Right side of the screen are a number of values identified by abbreviations
defined below:
POS
Rotary Valve position
STEP
Rotary Valve position required
VR1-6
Binary Rotary Valve position
TP
Retort direct and average pressure
OVERFILL
Overfill sensor status
VACUUM
Retort pressure management
GRADIENT
Time necessary to make vacuum and pressure in the Retort
ALARM
Running alarm code
RETORT
Number of the current contents of the retort
/D
Number of the last reagent in the Retort (0 If Purge Cycle Run)
SETP RET
Retort temperature set point
T DIR RET
Retort direct temperature
T MED RET
Retort average temperature (on 20 direct reads)
T MAX RET
Retort maximum temperature raised after the last reset
SETP WAX
Paraffin Chamber temperature set point
T DIR WAX
Paraffin Chamber direct temperature
T MED WAX
Paraffin Chamber average temperature (on 20 direct reads)
T MAX WAX
Paraffin Chamber maximum temperature raised after the last reset
1 2 . 3
–
S
E R V I C E
A
S S I S T A N C E
Before calling TBS
®
Product Service, please collect the following information:
ü
Type of instrument
ü
Serial number
ü
Software release
ü
The number of any alarm that has occurred
ü
A print of the Alarm Archive (if the printer is available)
Please provide this information to Product Service upon contact. Our telephone
number is
919-384-9393.
The department is staffed
8:00 AM to 6:00 PM EST or
DST
.