51
CHAPTER
11
–
ALARMS
1 1 . 1
–
M
A N A G I N G
A
L A R M S
There are two types of alarms:
N o n - B l o c k i n g A l a r m s
Alarms 1 through 9 are NON-BLOCKING alarms or WARNINGS that
do not
interrupt the program.
The event that caused the alarm is not critical for completion
of the process, and any program stoppage could produce worse results than
continuing.
With NON-BLOCKING alarms, the program will continue and the screen will display
the alarm number and a brief explanation of the alarm. In this situation, it is
NOT
desirable to interrupt the process. Upon completion, the cause of the alarm can be
determined and addressed by the operator.
Procedures listed below for each type of alarm should be followed to correct the
cause of the problem. If following all of the recommended steps to stop the alarm
from reoccurring does not help and/or if the problem persists for several processes,
the TBS
®
service department should be contacted.
B l o c k i n g A l a r m s
Alarms 11 through 53 are blocking alarms
.
These are true alarms and
do interrupt
process
, as the cause of the alarm makes it impossible to proceed. Continuing the
program could cause more serious damage than stopping it.
In the event of a BLOCKING alarm, follow the instructions displayed on the screen:
1) Stop the acoustic alarm.
2) Try to empty the retort.
3) Remove the samples.
4) Reset the instrument.
After the reset, the instrument is again ready to run a program. The alarm could
have been simply an error in closing the retort lid or something else easy to resolve.
Before proceeding, please check the following:
ü
Remove all samples.
ü
Check that the retort does not contain any reagent residuals.
ü
Check reagent and paraffin bottle levels.
ü
Check that the bottles are correctly placed in their housing slots.
ü
Check that the paraffin waxes are melted and, if possible, verify their
temperature.
ü
Check and clean the retort lid gasket.
ü
Execute a Purge Cycle to verify the functioning of the instrument.