The processing results will be the ultimate indicator as to whether the
RMS
settings
are ideal. To correct inadequate infiltration replace all the reagents, use the counters
reset functions to set the group counters to zero and decrease the process/cassette
number limits. Alternately, excellent results may indicate that the limits could be
increased
slightly
, extending the life of the reagents.
The
RMS Instructional Screen
indicates what action to take.
Similar to Single Bottle mode, at the end of a process the
RMS
will prompt the user
with graphics and text to replace a reagent bottle from one or more groups. This
task may be postponed to a later time by pressing the
ESC
key in the same manner
mentioned above, and, may be executed at the end of the next process or accessed
from the Main Menu.
When there are multiple concentrations belonging to a group such as the
DEHYD groups, the
highest
concentration solution should be utilized to
replace the reagent in the bottle.