45
•
Completion time
cannot
be set less than current time plus the time to
complete process. For example, if you start a 12-hour program at 9:00pm and
desire an end time of 7:30am, there is not enough time to complete the
process without cutting its time short. The computer will, therefore,
automatically adjust the Process End time so that the 12-hour processing
schedule is maintained, in this case, by finishing at 9.00am the following
morning.
It is highly recommended that the user verify the time, day and date
of process completion before starting a program. If it is set
incorrectly, the user’s only options are to wait for process
completion or to remove partially processed tissues from the retort,
reprogram and begin a new process. Processing programs cannot
be edited once they are begun.
M i d w e e k h o l i d a y s
There are two convenient solutions for handling midweek holidays:
•
Create a third standard program, named for example “HOLIDAYS”, where you
can set a different completion time to meet the particulars of your schedule.
•
Use the Daily setting, but be certain to reset the completion time/date before
utilizing the program again.
S t a r t i n g f r o m s t e p o t h e r t h a n t h e f i r s t
Use the
↑
and
↓
arrows. Note that the bar will move from the first step to every valid
step (every step is valid if its value is greater than 0). It is possible to select a
starting step different from the first step in a particular program.
•
This feature could be used, for example, on partially processed tissue
samples to restart the processing where it was interrupted.
R u n t i m e T e s t A b o r t i n g R u n
It is possible to practice setting program parameters, and even to launch a program,
because during the first phase of RUNTIME
TEST
, you can stop the program simply
by pressing the
ESC
key, as the instrument has not yet filled the Retort. Be certain
to re-enter the desired values before starting any programs.