23
7 . 6
–
H
O U R A N D
D
A T E
Select
HOUR
and DATE
from the S
etup Menu
.
The
Hour and Date Screen
:
With the arrow keys, point the cursor at what is to be changed, then increase or
decrease its values with the
+
and
-
keys.
Press
ENTER
to save changes and to go back to the previous menu.
Press
ESC
to abort changes and to go back to the previous menu.
7 . 7
–
P
U R G E
R
E A G E N T S
L
I M I T S
S
E T T I N G
Access these settings via
SETUP/RMS
. The reagents for the Purge Cycle are
managed by the
RMS
. It is necessary that the
RMS
be active in order to receive
reminders to substitute the Purge reagents every 3 to 9 processes. The
recommended numbers of processes which can be performed before the substitution
of the Purge reagents are as follows:
•
Cleaning only the retort: 9 processes or less
•
Cleaning the retort and the baskets: 5 processes or less
These recommendations are valid for the correct use of the instrument. However,
they are contingent upon the manual removal of as much residual paraffin as
possible (from the lid and retort) before beginning the Purge cycle.