50
Purge Cycle
procedure
:
The bottles containing the Purge Cycle reagents are labeled with P1 and P2 and
must contain: (the first from the left, top row) (P1) xylene (or substitutes), the second
(P2) 80% or 100% alcohol.
The Purge Cycle reagent (if not managed by the RMS) must be changed
at least every 5 completed paraffin processes
.
Take care not to leave fabric or paper residue (normally used in cleaning
procedures) as the material can prevent the complete closing of the main
valve membrane with subsequent damage or alarms. Pieces of fabric
can damage the rotating valve, causing reagent leaks and cross-over
contamination between reagents.
The last Purge step is the drying of the Retort of any alcohol residue. This step is
not critical and can be stopped at any time without damaging the instrument.