LIFEPAK
®
20
Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 38 of 61
PIP - Standard Paddles Defibrillator a Positive R-Wave Test
To perform A positive R-wave test:
1. Set the Impulse 7000DP to a 1mV, Normal Sinus, 60 BPM.
2. Set the device ECG SIZE to 1.0.
3. Set the
LEAD
selection to
PADDLES
.
4. Print 10 seconds of paddles ECG recorded on printer
paper.
5. Turn the printer OFF.
6. Confirm the R-Wave polarity on the paper matches image
as shown in Figure 1.24.
Figure 1.24: A Positive R-Wave