LIFEPAK
®
20
Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 33 of 61
PIP- Standard Paddles Defibrillator Sync Test at 2J
Perform this test only if the device is equipped with the standard paddles option.
WARNING
SHOCK HAZARD
Electrical energy is discharged during this procedure. Do not allow the electrodes to contact any person or conductive surface
except as described below.
Note:
Ensure that the Standard Paddles is connected between the device and the Impulse 7000 DP, using the appropriate adapters.
Note:
Verify that the metal surface of the paddles are free of burn or arc marks and are free of pits, scratches, insulting films,
contaminants or raised nicks that can be felt with the finger tip.
Note:
To avoid damage to the Defibrillator Analyzer, do not apply any defibrillator pulses to the pacer inputs of the analyzer.