Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 53 of 61
To set up the Safety Analyzer to measure Direct Applied Part Leakage:
1. Press the
A button on the Safety Analyzer.
2. Press the F4/MORE button.
3. Press the UP/DOWN arrows to select the appropriate A.P groups as shown in
Figure 1.29.
4. Press F1/SELECT then F1/Direct A.P.
Figure 1.29: Direct Applied Part
Leakage Setup