LIFEPAK
®
20
Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 34 of 61
To perform Standard Paddles Sync Test at 2J
1. Establish the setup as shown in Figure 1.21.
2. Turn the device
ON
.
3. Set the device:
MANUAL MODE
,
ECG Size to 1.0
,
LEAD
selection to
LEAD II
.
4. Set the Defibrillator Analyzer to measure
SYNC
.
5. Press the
SYNC
button
ON
and select
LOCAL
, if Remote Sync is set to on.
6. Verify the Sync LED turns on and R-wave markers appear on the ECG
waveform.
7. Press the
ENERGY SELECT
button to select 2J.
8. Press the
CHARGE
button on Standard Paddles and wait for the device to
reach full charge. Then push both
SHOCK
buttons on Standard Paddles to
discharge the device.
9. Verify the defibrillator analyzer measures a sync R-wave is within 1.0 ms to
60ms.
10. Continue to the next test with this setup in place.
Figure 1.21: Standard Paddles Defibrillator
Energy Test Setup