Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 41 of 61
– Pacer Leads-Off Detection Test
To perform Pacer Leads-Off detection:
1. Establish the setup as shown in Figure 1.26.
Note:
Can use any type of ECG Lead cable in the setup.
2. Turn the device
ON
.
3. Set the Impulse 7000DP to measure Pacer
4. Press
PACER
button, press F1/Brand to Physio Control, press F2/Input to Defib, and
press F3/Load to 50 ohms.
5. Press
PACER
on the device.
6. Verify the PACER LED is on and the Pacer overlay appears.
7. Disconnect one of the test post adapter snaps from the Impulse 7000DP.
8. Verify the Pacing/Connect Electrodes overlay appears, accompanied by an audible
alarm.
9. Reconnect the test post adapter snap.
10. Verify the Pacing/Connect Electrodes overlay disappears and the alarm stops.
Figure 1.26: Pacer Test Setup