LIFEPAK
®
20
Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 39 of 61
PIP - Standard Paddles Sync Remote Test
Note:
Perform this test if Standard Paddles is used. Otherwise, skip to the next test.
To perform Standard Paddles Sync Remote Test:
1. Establish the setup as shown in Figure 1.25.
2. Turn on the Remote Sync Test Pulse Generator.
Note:
A function generator set to provide a pulse train 5 Vp-p (0-5 V), 5 to
200 ms wide, 120 PPM (2 Hz) can be used alternatively to the Remote
Sync Test Pulse Generator.
3. On the Manual Mode setup page; set the device to Remote Sync
ON
.
4. Turn the device
OFF
and then turn the device
ON
.
5. Set the device:
Manual mode
, lead selection to
PADDLES
.
6. Press the
SYNC
button on the device.
7. On the Sync mode screen, select
REMOTE
.
8. Verify the SYNC LED is flashing.
9. Charge the device to 200 J.
10. Upon reaching full charge, press the
SHOCK
button to discharge the device.
11. Verify the device displays
"ENERGY DELIVERED"
screen message (for SW-
20 version or below) or switches out of remote sync mode (for SW-26 or
above).
12. On the Manual Mode setup page, set the device to Remote Sync
OFF
.
13. Turn the device
OFF
Figure 1.25: Standard Paddles Sync
Remote Test Setup