Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 31 of 61
PIP - Quik-Combo Sync Remote Test
Note:
Perform this test if Quik-Combo cable is used. Otherwise, skip to the next test.
To perform Quik-Combo Sync Remote Test:
1. Establish the setup as shown in Figure 1.19.
2. Turn on the Remote Sync Test Pulse Generator.
Note:
A function generator set to provide a pulse train 5 Vp-p (0-
5 V), 5 to 200 ms wide, 120 PPM (2 Hz) can be used
alternatively to the Remote Sync Test Pulse Generator.
3. On the Manual Mode setup page; set the device to Remote Sync
ON.
4. Turn the device
OFF
and then turn the device
ON
.
5. Set the device:
Manual mode
, lead selection to
PADDLES
.
6. Press the
SYNC
button on the device.
7. On the Sync mode screen, select
REMOTE
.
8. Verify the
SYNC LED
is flashing.
9. Charge the device to 200 J.
10. Upon reaching full charge, press the
SHOCK
button to discharge
the device.
11. Verify the device displays
"ENERGY DELIVERED"
screen
message (for SW-20 version or below) or switches out of remote
sync mode (for SW-26 or above).
12. On the Manual Mode setup page, set the device to Remote Sync
OFF
.
13. Turn the device
OFF
.
Figure 1.19: Quik-Combo
Sync Remote Test Setup