Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 43 of 61
To perform the pacer pulse width test:
1. Set pacer rate on the device at 60 PPM.
2. Press
CURRENT
on the device, and select a pacer current of 200 mA.
3. Verify the measured pacer pulse width is between 19.0 and 21.0 ms.
Note:
Perform TCP - Pacer Self-Calibration if the peak pacer current falls outside the acceptable output range.