PIP - QUIK-COMBO Defibrillator Sync Test at 2J
To perform QUIK- COMBO Defibrillator Sync Test at 2J:
Note:
Perform this test if Quik-Combo cable is used. Otherwise, skip to the next test
.
To perform Quik-Combo Defibrillator Sync Test at 2 J:
1. Establish the setup as shown in Figure 1.15.
Note:
Ensure proper test setup connections to the Defibrillator Analyzer. To avoid
damage to the Analyzer or defibrillator, do NOT apply defibrillator pulses to the pacer
inputs of the analyzer.
2. Turn the device
ON.
3. Set the device:
Manual
mode,
ECG Size
to 1.0,
LEAD
selection to
Lead II
.
4. Set the Defibrillator Analyzer to measure
SYNC
.
5. Press the
SYNC
button
ON
and select
LOCAL
, if Remote Sync is set to on.
6. Verify the Sync LED turns on and R-wave markers appear on the ECG waveform.
7. Press the
ENERGY SELECT
button to select 2J.
8. Press the
CHARGE
button and wait for the device to reach full charge. Then push
the
SHOCK
button to discharge the device.
9. Verify the defibrillator analyzer measures a sync R-wave is within 1.0 ms to 60ms.
10. Continue to the next test with this setup in place
.
Figure 1.15: Quik-Combo
Defibrillator Delivered Energy Setup