LIFEPAK
®
20
Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 11 of 61
It is recommended that the device be set up for Manual mode when performing the PIP.
Note:
If you do not wish to change the setup for a device configured with manual access restrictions, it may be necessary to use the
reserved technician passcode of 5433 to gain access to Manual mode.
Note:
Be sure to make note of the customer settings to restore the device to the user-selected
MANUAL ACCESS
configuration at the completion of this PIP.
To perform the device for Manual mode access:
1. Access the Setup mode as follows:
a. Press and hold
OPTIONS
and
EVENT
, and then turn the device
ON
.
b. When the Setup mode passcode prompt appears enter 5433.
2. Select
MANUAL MODE
in the Setup menu.
3. In the Setup/Manual Mode submenu, set the Manual Access selection to
Manual/Direct
as shown in Figure 1.2.
4. Turn the device
OFF
by pressing
ON
for two seconds.
Figure 1.2: Manual Mode Access